Fiducial-based imaging is often used in IGRT. Traditional gold fiducial marker often has substantial reconstruction artifacts. These artifacts Result in poor image quality of DRR for online kV-to-DRR matching. This study evaluated the image quality of PEEK in DRR in static and moving phantom.
CT scan of the Gold and PEEK fiducial (both 1×3 mm) was acquired in a 22 cm cylindrical phantom filled with water. Image artifacts was evaluated with maximum CT value deviated from water due to artifacts; volume of artifacts in 10×10 cm in the center slice; maximum length of streak artifacts from the fiducial. DRR resolution were measured using FWHM and FWTM. 4DCT of PEEK fiducial was acquired with the phantom moving sinusoidally in superior-inferior direction. Motion artifacts were assessed for various 4D phase angles.
The maximum CT value deviation was −174 for Gold and −24 for PEEK. The volume of artifacts in a 10x10 cm 3 mm slice was 0.369 for Gold and 0.074 cm3 for PEEK. The maximum length of streak artifact was 80mm for Gold and 7 mm for PEEK. PEEK in DRR, FWHM was close to actual (1.0 mm for Gold and 1.1 mm for PEEK). FWTM was 1.8 mm for Gold and 1.3 mm for PEEK in DRR. Barrel motion artifact of PEEK fiducial was noticeable for free-breathing scan. The apparent PEEK length due to residual motion was in close agreement with the calculated length (13 mm for 30–70 phase, 10 mm in 40–60 phase).
Streak artifacts on planning CT associated with use of gold fiducial can be significantly reduced by PEEK fiducial, while having adequate kV image contrast. DRR image resolution at FWTM was improved from 1.8 mm to 1.3 mm. Because of this improvement, we have been routinely use PEEK for liver IGRT.
Read Full Story from Medical Physics: Most Recent Articles http://scitation.aip.org/content/aapm/journal/medphys/42/6/10.1118/1.4924125?TRACK=RSS
This article by S Shen, R Jacob, R Popple, J Duan, X Wu, R Cardan and I Brezovich originally appeared on scitation.aip.org on June 26, 2015 at 06:59PM